Skip to main content
Fig. 8 | EPJ Techniques and Instrumentation

Fig. 8

From: Calibration of a multi-anvil high-pressure apparatus to simulate planetary interior conditions

Fig. 8

a The temperature distribution in the assembly, as modelled with the numerical model of Hernlund et al. [43]. Constant colour areas are indicative of sample (black), graphite (grey), Al2O3 thermocouple tubing (green), MgO (cyan), Zr2O3 (yellow) and WC cubes (darkish cyan). Black contours lines represent isotherms separated by 50 degree intervals, with the 1200 °C isotherm located closest to the sample (at z = 1.5 mm and r = 0 mm). b The modelled temperature profile along the assembly’s axis (the r = 0 line)