Fast neutron sensitivity for 3He detectors and comparison with Boron-10 based neutron detectors

Abstract

The 3He-based neutron detectors are no longer the default solution for neutron scattering applications. Both the inability of fulfilling the requirements in performance, needed for the new instruments, and the shortage of 3He, drove a series of research programs aiming to find new technologies for neutron detection. The characteristics of the new detector technologies have been extensively tested to prove their effectiveness with respect to the state-of-the-art technology.Among these, the background rejection capability is crucial to determine. The signal-to-background ratio is strongly related to the performance figure-of-merit for most instruments. These are designed to exploit the high flux expected from the new high intensity neutron sources. Therefore, an inadequate background rejection could significantly affect the measurements, leading to detector saturation and misleading events. This is of particular importance for the kind of techniques in which the signals are rather weak.For the first time, the sensitivity of 3He detectors to fast neutrons, up to En=10 MeV, has been estimated. Two independent measurements are presented: a direct calculation based on a subtraction method used to disentangle the thermal and the fast neutron contribution, while a further evidence is calculated indirectly through a comparison with the recently published data from a 10B-based detector. Both investigations give a characterization on the order of magnitude for the sensitivity. A set of simulations is presented as well in order to support and to validate the results of the measurements. A sensitivity of 4·10−3 is observed from the data. This is two orders of magnitude higher than that previously observed in 10B-based detectors.

Introduction

The increasing complexity in science investigations, driven by technological advances, is reflected in the studies of neutron scattering science which enforces a diversification and an improvement of experimental tools, from the instrument design to the detector performance. Complexity means study of realistic and ever more complex heterogeneous samples within extreme or natural environments, and investigation of multiple physical properties of materials [1]. Therefore, the instruments must be flexible, permitting exchanges between brightness and constrained phase-space: in turn implying better resolution, optimization of signal-to-background ratio, and the use of polarized neutrons when necessary. The increasing performance demands drive for development of neutron detector technologies; not only the enhanced detector response is of a crucial importance, but the background rejection plays a significant role as well.

The new generation, high-intensity neutron sources like the Spallation Neutron Source [2] (Oak Ridge National Laboratory, TN, USA), the Japan Spallation Source [3] (Tokai-mura, Japan), and the European Spallation Source [1, 4, 5] (ESS, Lund, Sweden), presently under construction, where an intensity increase of at least one order of magnitude further is expected [68], represents a challenge due to the increasing demands on neutron detectors. The trends in neutron scattering applications point out the importance of exploiting long-wavelength neutrons with energies in the range of 0.1 meV to 80 meV (30Å- 1Å). The high brightness, foreseen at ESS, in this energy range will allow faster measurements, increased use of polarized neutrons, investigation of smaller samples and detection of weaker signals. The actual limitations are not only set by the available neutron flux, but also by the detector response.

The 3He-based technology has been predominant for thermal neutrons detection. Both the availability [9] and the requirements of higher performance, e.g., of counting rate capability, spatial resolution and background sensitivity, are the reasons why a number of research programs are now aiming to find technologies that would replace 3He [1012]. A promising technique is based on solid converter layers (10B, Gd) and either gas proportional counter or solid state material as the sensing medium. Some examples that employ 10B are: the Multi-Grid [1316], the Multi-Blade [1719], the Jalousie detector [20], BandGEM [21, 22], the Boron-coated straw-tubes [23] and CASCADE [24]. Other technologies are based on Gd, e.g., the Gd-GEM [25, 26], and a whole branch dedicated to the development of solid state neutron detectors coupled with Gadolinium as a converter layer [2730]. This research phase involves the investigation of all the characteristics of a prototype, not only to prove the effectiveness of the new technology, but to demonstrate their relative performance to 3He detectors.

The signal-to-background ratio is one of the features that strongly affect the figure-of-merit for most neutron scattering instruments, especially when the detected signals are rather weak and the background discrimination is crucial to disentangle the two effects. The signal is defined by the source brightness and phase space acceptance, and a better background rejection can improve this figure-of-merit, which is a function of a more complex combination of instrument and detector geometry and parameters, leading to significant impact on instrument’s operation, especially at the new high intensity sources.

A previous investigation of the fast neutron sensitivity has been performed on a 10B-based thermal neutron detector [31], the Multi-Blade detector [17, 18]. A sensitivity on the order of 10−5 has been measured, about 103 times higher than the gamma sensitivity observed with the same detector [31, 32]. The fast neutrons make a significant contribution to the background, hence, it is important to study the detector response to it.

This paper investigates, for the first time, the fast neutrons sensitivity of 3He-based neutron detector. The discussion is presented in comparison with the Muti-Blade detector, whose response to both γ-ray [32] and fast neutrons [31] has been already studied. The fast neutron sensitivity has been calculated as a function of energy threshold. Moreover, a set of simulations has been performed to verify the subtraction method applied to the measurements. From a campaign of measurements perfomed at the CRISP [33] reflectometer at ISIS (Science & Technology Facilities Council in UK [34]), a direct comparison of the fast neutron sensitivity between an 3He-based detector and the Multi-Blade 10B-based detector [18] is presented. Both results are aimed at quantitative analysis on the order of magnitude; higher accuracy cannot be achieved with these kind of measurements.

Methods/Experimental

Set-up description and methodology

A set of measurements has been carried out at the Source Testing Facility of Lund University [35] with a 3He-tube, to investigate the fast neutron sensitivity of this class of devices. The tube used is a RS P4 0810 227, manufactured by General Electric (GE) Reuter Stokes [36]. It is made of stainless steel and it has an active length of 10 inch (250 mm), a radius of 1 inch (25 mm), a gas pressure of 10 bar and an efficiency of 96.4% (at 2.5Å).

Two fast neutron radioactive sources have been employed: an 241Am/Be [37] and a 238Pu/Be source [38]. The fast neutron energy range is definined by the emission of these sources, it extends up to approximately En=10 MeV. A measurement was also performed with a 60Co gamma source to have a comparison with a previous work on gamma sensitivity performed using the same 3He detector [39].

With the previous evaluation of sensitivity to fast neutron with the 10B-based detector, excellent separation of the fast neutron and thermal neutron contributions was possible by using identical detectors one containing the 10B4C layer, sensitive to both contributions, and a second detector without the converter layer, only sensitive to fast neutrons. However, it is not trivial to have a good separation between the different neutron contributions, when using the 3He-tube, because of the high efficiency to thermal neutrons.

Note that with the terms efficiency and sensitivity it is intended the same mathematical quantity. As defined in Eq. 4, it is a function of the detected events and the total incident events on the detector. We refer to efficiency in case of thermal neutron counts, whereas sensitivity is used to identify any other detected radiation, i.e. gamma and fast neutron contribution.

The measurements were performed in several configurations, a sketch of which is shown in Fig. 1. In configuration (FastN), Fig. 1a, the total flux of the source reaches the detector, i.e., fast and thermal contribution. In configuration (ThermN), Fig. 1b, the flux is partially thermalized through a polyethylene brick, while in (BackgN), Fig. 1c, a shielding of polyethylene, borated-polyethylene and lead is placed between the source and the detector, in order to reduce as much as possible the incoming neutrons both fast and thermal. The latter is a measurement of the background with the source in place. In order to reduce the background due to the scattering from the environment, the detector is placed in a shielding sleeve of boron powder of 4 cm radius, with a slit opening in the direction of the source. Moreover, borated polyethylene and lead slabs are placed at the sides along the source-detector distance. This, strongly reduce the scattering effects from the surrounding and the opening in the shielding sleeve identify the actual active area of the detector which allows the solid angle calculation.

Fig. 1
figure 1

Configurations used in the measurements with the 3He-tube: a the total flux reaches the detector, b flux thermalized through a polyethylene brick, c flux highly reduced through polyethylene, borated polyethylene and lead bricks, in order to perform a background measurement with the source

The detector-source distance is d = 250 mm, the depth of the polyethylene used in configuration (ThermN) is 100 mm, while in the case of (BackgN) the depths of polyethylene, borated polyethylene and lead are 40 mm, 25 mm and 2 mm, respectively. These values are used in the simulation to reproduce the actual set-up. The solid angle has been calculated on a rectangular surface of the detector (150x10) mm2 for the fixed distance d. Considering a point-like source, it results in a solid angle of 0.002 sr (0.2% of a sphere). It has been estimated that the measurement of the distance between the source and the sensitive region can lead to an uncertainty which is less than a factor two variation for a misplacement of a ±25 mm from the fixed location of the source [31]. The uncertainties in these measurements are dominated by this systematic effect, whereas the statistical uncertainties can be considered negligible as they depend on the counting statistic.

A subtraction method has been used to disentangle the thermal and fast contribution. Referring to Fig. 1, we define the flux reaching the detector in the three measurements as:

  1. 1.

    FastN=ϕfn+φtn

  2. 2.

    ThermN=φfn+ϕtn

  3. 3.

    BackN=φfn+φtn

Where ϕ define the main contribution of the incoming neutron flux, fn in the case of fast neutrons and tn for thermal neutrons; φ is the component of flux which is considered as background with respect to the main contribution. By subtracting BackN from the other two contributions, we obtain:

$$\begin{array}{*{20}l} \text{FastN} - \text{BackN} = \phi_{fn} + \varphi_{tn} - (\varphi_{fn} + \varphi_{tn}) = \phi_{fn} -\varphi_{fn} \end{array} $$
(1)
$$\begin{array}{*{20}l} \text{ThermN} - \text{BackN} = \varphi_{fn} + \phi_{tn} - (\varphi_{fn} + \varphi_{tn}) = \phi_{tn} - \varphi_{tn} \end{array} $$
(2)

The subtraction allows to separate the fast neutron and thermal neutron contributions, as shown in equations 1 and 2. We recall that this method give an order of magnitude accuracy and it is valid when the background contributions (φ) are of lower intensity than the main contributions (ϕ). The details are discussed in “Simulations” section and “Fast neutron sensitivity of 3He detector” section. A similar discrimination, based on a more complex analytical procedure using the Inverse Matrix Method can be found in [40]. The subtraction method has been applied on both simulations and measurements.

The normalization to the flux, number of neutrons per unit time and area, is obtained multiplying the activity of the source by the solid angle [31]. Note that, for (ThermN) only a fraction of thermalized flux reaches the detector, therefore we calculate the total and thermalized flux from the simulations. The calculated flux, C, is the integral of counts in a selected incident neutron energy range. Note that the solid angle is included in this calculation for the simulation.

$$ C = \int_{E_{1}}^{E_{2}} C(x) \, \mathrm{d}x $$
(3)

Where C(x) is the energy distribution of the counts, while E1 and E2 define the energy range of integration. This range varies for each configuration: in the case of (FastN) the energy limit is set by the detectable energy, E2≈1 MeV. For (ThermN) it is limited by the thermal neutron energy, for this study we consider E2=0.050 eV.

Once the normalization is calculated, we can derive the expression for the sensitivity as the integral of the normalized counts, f(x), above a certain energy threshold:

$$ \epsilon = \int_{E_{th}}^{E_{f}} f(x) \, \mathrm{d}x $$
(4)

Where Ef is the maximum detected energy and Eth is the selected energy threshold. A similar approach using threshold discrimination can be found in literature for B-10 loaded scintillator detectors [4143] compared with 3He detectors, including pulse shape analysis.

Results/Discussion

Simulations

The neutron reactions involved in the case of a 3He counter are four:

  1. 1.

    3He recoil 3He+n3He+n (elastic scattering),

  2. 2.

    (n,p) 3He+np+T+764 keV,

  3. 3.

    (n,d) 3He+n2H+2H−3.27MeV,

  4. 4.

    γ-rays.

The contribution of the (n,d) reaction is negligible because of the high threshold energy (>4 MeV) and the small cross section for this process. Below 1 MeV the (n,p) reaction has the highest cross section, i.e., contributes mostly to the energy deposition in a 3He tube. The cross section for this process is proportional to 1/v, where v is the velocity of the neutron. It is used in the detection of thermal neutrons, because the cross section increases as the kinetic energy of the neutron decreases. Considering the energy region above 1 MeV, the 3He recoil is the predominant process. In Fig. 2 the cross sections for the different mechanisms between 10−5 and 107 eV are shown. The vertical line indicates the separation at 1 MeV. The data refers to the NIST web database [44].

Fig. 2
figure 2

Neutron cross section of 3He in the energy range 0-10 MeV for the 4 reactions: 3He recoil, (n,p), (n,d) and γ-rays. The total cross section is shown as well. Data from [44]

To support the experimental analysis a Geant4 [4547] simulation model is built within the ESS Detector Group simulation framework [48]. The geometry matches the experimental set-up for all three configurations, as seen in Fig. 3. The counting gas inside the tube is pure 3He, while the material of the steel vessel is represented by γ-iron with the help of the NXSG4 library [49], which treats the thermal neutron transport. The thickness of the steel is 0.4 cm. The wireframe volume appearing in the geometry represents the shielding used in the set-up. Every neutron incident on it is killed. The yellow volume is the window that shapes the active area of the detector, which reproduce the effect of the slit opening of the shielding sleeve surrounding the 3He tube. In order to reproduce a more realistic geometry of the experiment and to evaluate the effects of the background at the detector, further simulations have been performed including concrete walls, wooden table and the side shielding. Including these items in the simulations increases the difference between the configurations (FastN - BackgN) and (ThermN - BackgN) no more than 4%. Being the aim of this manuscript to give an order of magnitude quantification for the fast neutron sensitivity, the simplified geometry (Fig. 3) has being used for the calculation.

Fig. 3
figure 3

Simulated geometry of configurations (FastN) (a), (ThermN) (b) with the polyethylene brick and (BackgN) (c) with the polyethylene brick and the alternate lead and borated polyethylene slabs in black and dark blue respectively. The detector appears in gray. The wireframe volume acts as the neutron shielding used in the experimental set-up. The yellow window lets the neutrons into the detector active area. Generated neutrons are depicted in green

The neutron generator approximates the Pu/Be source energy distribution but instead of a 4 π emission, a conical pattern is adopted for better exploitation of statistics. A custom physics list is used to enable correct thermalization in the polyethylene [50].

The energy released in a 3He detector, with the characteristics described in the previous section, is shown for the configuration (FastN) in Fig. 4 and for configuration (ThermN) in Fig. 5. In both cases, the contribution for each of products of the reactions are shown in the energy range below 900 keV, which is the detectable range. As expected, the elastic scattering, 3He recoil, contributes mostly to the total Pulse Height Spectrum (PHS) when the fast neutron energies (1-10 MeV in case of the used sources) are taken into account in the simulation, as shown in Fig. 6. On the other hand, for the thermalized flux, the products of the (n,p) process, proton and tritium, give the highest contribution to the energy distribution.

Fig. 4
figure 4

Simulated energy deposition for the reaction products in configuration (FastN)

Fig. 5
figure 5

Simulated energy deposition for the reaction products in configuration (ThermN)

Fig. 6
figure 6

Simulated total energy deposition for configurations (FastN), (ThermN) and (BackgN) in the full energy emission range of the Pu/Be source

The full spectrum of the simulated energy deposition for the three simulation set-ups is shown in Fig. 7. With the shielding used in configuration (BackgN) the thermal peak is about 20% of the peak recorder in (ThermN) configuration. From the simulations we calculate φtn=2% ϕtn, which are the neutrons that thermalizing cross the shielding and reach the detector, and φfn=17% ϕfn for the (FastN) configuration. We can assume that with the (BackgN) measurement we identify mostly the neutrons that scatter and/or thermalize from indirect source-detector line of view, i.e. scattering from the environment around the detector. Therefore, this contribution is present in both (FastN) and (ThermN) measurements and can be removed applying the subtraction method. This assumption is valid to lead to the evaluation of the order of magnitude of the fast neutron sensitivity which is the aim of the work presented in this manuscript.

Fig. 7
figure 7

Simulated total energy deposition for configurations (FastN), (ThermN) and (BackgN)

Note that no normalization is taken into account yet. The estimation of the flux of neutrons reaching the active area of the detector is fundamental to calculate the flux normalization in the different sets of measurements.

The energy distribution of incident neutrons in the detector active area are shown in Fig. 8, both for the (FastN) and (ThermN) configurations. We estimate that the thermalized flux (En<0.1 eV) passing through a 100 mm layer of polyethylene is ≈8% of the total neutron counts that reach the detector for the full energy range considered for the Pu/Be source. This value is in agreement with [51].

Fig. 8
figure 8

a Incident neutron energy in the detector active area for configurations (FastN) and (ThermN). The energy range 0-10 MeV is the 238Pu/Be source emission spectrum. b The counts in 0<En<0.4 eV are shown. The vertical lines represent the cut at the E2=1 MeV and E2=0.050 eV used to calculate the fast and thermal flux respectively

From the simulation we can derive the thermalized flux obtained in configuration (ThermN) sketched in Fig. 3b as described in the previous section following equation 3. The integrated counts, ≈3.5·105, have been calculated in an energy range below E2=0.050 eV, as shown in panel (b) of Fig. 8. In the case of (FastN), Fig. 3a, the flux has been calculated based on equation 3 up to E2=1 MeV, the cut is shown in Fig. 8, panel (a) (vertical line). This is about 1 order of magnitude higher than the thermalized flux. The ratio between the two calculated fluxes will be used to normalize the measured pulse height spectra presented in “Fast neutron sensitivity of 3He detector” section. As described in the previous section for the measurements, the flux is calculated by multiplying the activity and the solid angle for the fast neutron configuration. The ratio calculated from the simulations (≈10%) is needed to normalize the (ThermN) measurement. Indeed, this information cannot be extracted from the data.

Before applying the normalizations to the PHS, we apply the subtraction method to disentangle the fast and the thermal neutrons. After the subtraction, we proceed with the normalization.

The efficiency of the detector can be calculated by integrating the counts of the normalized PHS above a certain energy threshold as described in equation 4. In Fig. 9 the sensitivity both to thermal and fast neutrons is shown. For a threshold of 100 keV, which is usually chosen to discriminate the background [31, 32], a fast neutron sensitivity of about 3.3·10−3 is obtained and an efficiency of 94% for thermal neutrons, in agreement with the specification of the detector. Moreover, the sensitivity is in the order of 10−3 up to a threshold of Eth=400 keV.

Fig. 9
figure 9

Fast neutron sensitivity of the 3He-detector as a function of an energy threshold Eth=100 keV. The threshold value has been taken in agreement to the threshold used for background discrimination based on previous works [31, 32]

Fast neutron sensitivity of 3He detector

As described in section 3, several sources have been used: Am/Be and Pu/Be to measure the fast neutrons, the Pu/Be source has been employed to perform the thermal neutron measurement as well. A Co-60 source was used to measure the γ-ray sensitivity, as a comparison with previous work [39]. Measurements of the environmental background, without any source, were performed as well. The PHS, normalized over time, are depicted in Fig. 10; the centre of the peak corresponding to an energy of 764 keV is used to convert the X-axis from ADC values to energy. The same method is applied to each PHS reported below.

Fig. 10
figure 10

Pulse Height Spectrum (PHS) normalized over time for different sources: Am/Be and Pu/Be for the fast neutrons and Co-60 for γ-rays, and the environmental background

The Am/Be source has a larger γ emission with respect to the Pu/Be source [31, 38]. The effect is clearly visible, indeed, the energy released by γ-rays spreads up to about 200 keV (green curve). Below this region (<100 keV) the PHS measured with the Am/Be source is about one order of magnitude more intense than that recorded with the Pu/Be source. The lower γ-rays emission and the highest activity are the reasons why the presented measurements were carried out with the Pu/Be source.

In Fig. 11 the PHS measured in the configurations (FastN), (ThermN) and (BackgN) are shown. The background counts due to the environment, Fig. 10, have been subtracted from this data sets. The spectra are normalized by time and solid angle.

Fig. 11
figure 11

Measured PHS normalized by time and solid angle, obtained in configuration (FastN), (ThermN) and (BackgN)

In configuration (BackgN) the reaction product peak (764 keV) is only one order of magnitude less intense than in case of the thermalized flux configuration (ThermN). This value is in agreement with the simulation, see Fig. 7. The same peak is detected in the (FastN) measurement, which confirms the difficulty to decouple the fast and the thermal neutron contributions. The difference of the integrated peak in (FastN) and (BackgN) configurations is approximately 20%, which is in agreement with the ratio of total incoming neutrons calculated in the simulations. Therefore, the peaks are normalized by this difference before applying the subtraction method needed to disentangle the two outputs. From the simulation we know that φtn=2% ϕtn, hence we normalize according to this ratio the integral of the thermal peak obtained for the (BackgN) and the (ThermN) measurement, before applying the subtraction.

Based on the previous discussion, the flux in (BackgN) measurement can be considered as a background component, both due to the thermalized component passing after the shielding and fast neutrons that scatter around the detector thermalizing at the detector. Thus, this contribution can be subtracted in both scenarios. The resulting PHS are depicted in Fig. 12. The fast neutron spectrum is, as expected from the theory, mostly flat after applying the subtraction method.

Fig. 12
figure 12

PHS of thermal neutrons (blue) and fast neutrons (red). The (BackgN) measurement has been subtracted in the case of (FastN), in order to ensure that the spectrum refers to fast neutron events

Note that, a slight shape of the thermal peak is still distiguishable, but it is the best discrimination achievable in this measurements. When the energy range of the incoming source is in the fast neutron range, up to En=10 MeV in the case under study, the pulse height distribution for the 3He recoil reaction is continuous from E=0 to (3/4)En (Fig. 6), while for the (n,p) reaction the pulse height spectrum has a peak at a value proportional to En+Q and a continuous wall effect up to En+Q, in case a monochromatic energy is considered [40]. Since the pulse height spectrum is shown for energies below 900 keV a continuous distribution is expected. This result is in agreement with the simulated energy deposition shown in Fig. 7. Therefore, the subtraction method can be considered as a useful tool to disentangle thermal and fast neutron events.

The PHS in Fig. 12 are normalized by time and flux. The latter has been derived from the simulation. As described in section 3, we calculated the integrated flux after the thermalization through the polyethylene brick, the total incoming neutron flux, and the same for the fast neutrons, based on equation 3. For the measurement, the normalization is obtained from the activity of the source and the solid angle in the case of fast neutrons. The calculated ratio, ≈10%, between the thermal and fast flux from the simulation is used to normalize the PHS in (ThermN) configuration.

In Fig. 13 is shown the sensitivity for configuration (FastN) and (ThermN), calculated referring to equation 4. The Eth=100 keV threshold, red vertical line, is taken according to the simulation and the background discrimination based on previous works with some similar characterizations [31, 32]. A sensitivity of ≈4·10−3 is observed. The same order of magnitude is obtained both for measurements and simulations.

Fig. 13
figure 13

Fast neutron sensitivity of the 3He-detector as a function of the applied threshold. For a software threshold of 100 keV a fast neutron sensitivity on the order of 10−3 is obtained

Comparison of fast neutron sensitivity between 3He-based and Multi-Blade B10-based detectors

Background study from the test at the CRISP reflectometer at ISIS

The measurements presented in this section refer to a test performed at the CRISP reflectometer [33] at ISIS, with a Boron-10-based gaseous detector, the Multi-Blade [32], which is under development at the European Spallation Source, ESS [1, 4]. The Multi-Blade is a stack of units, each is a Multi Wire Proportional Chamber operated with a continuous gas flow of Ar/CO2. One module consists of a flat substrate (a blade) coated with a 10B4C neutron converter layer inclined at a grazing angle of 5 degrees with respect to the incoming neutron beam.

The full description of the experiment and the detector can be found in [18, 19]. A background study has been performed on a particular set of measurements, and a direct comparison between the 3He-tube installed at the instrument and the Multi-Blade detector employed for the tests, is reported in this paper.

For those measurements, the two detectors were placed in the direct beam with the chopper working at half of the nominal speed, i.e., 25 Hz instead of 50 Hz. Choppers are rotating disk with at least one aperture, which allow only neutrons of a certain velocity (energy) pass through it. When the neutron beam hits the chopper thermal neutrons are absorbed, γ-rays attenuated, but a fast neutron component can still pass through the chopper and reach the detector.

The CRISP reflectometer uses a broad band neutron time-of-flight method for determining the wavelength, λ, in a range of 0.5-6.5 Å when operating at the source frequency of 50 Hz, and can be extened up to a maximum of 13 Å, if operated at 25 Hz [33]. The larger wavelengths, lower neutron energies (cold range), will be exploited at the new high intensity source ESS. In this range, the efficiency of the Multi-Blade is comparable with that of the 3He detectors [18]. Moreover, at high q where the reflectivity intensity is very low, down to 10−6 at the current instruments, the signal-to-background ratio is crucial to perform an adequate data analysis.

As the instrument is working at half of the nominal speed, every second pulse the neutron beam find the chopper closed to let the longer wavelengths of the previous pulse to reach the sample, together with this, an intense prompt pulse is generated by the scattering between the neutrons and the chopper material. A sketch of the spectrum in a time window of T=200 ms is shown in Fig. 14. There are three possible scenarios: the chopper is in phase with the proton pulse, thus the beam passes through it (O in the picture), the chopper is not in phase with the frequency of the proton pulse (C in the picture), so the neutron beam hits the chopper producing a high background spike. Furthermore, every five pulses, one pulse is sent to the Target Station 2 (TS2 in picture), a negligible background is expected in this case, because there is no beam impinging the chopper disc.

Fig. 14
figure 14

Sketch of the arrival signals in a window of 200 ms, when the chopper works at 25 Hz (CRISP). The black identifies the signal passing throw the chopper (O = open), the red colour is used when the neutron beam hits the chopper without passing through (C = close). In this case a strong spike is expected at the detector. One every five pulses is sent to the Target Station 2 (TS2), blue in figure. A low background can be detected

The detector response can be analysed in different time windows of this spectrum. An indirect estimation of the fast neutron sensitivity is obtained focusing the analysis on the background contributions when the chopper is closed (C) in Fig. 14. A prompt pulse is expected in the case the chopper is open as well. All these events are mainly due to γ-rays, fast neutrons, environmental neutron counts (thermal and epithermal) and spurious scattering.

We define three different regions of studying: the full spectrum t=0−200 ms (T), the thermal spectrum region (O) in t=120−160 ms and the background peak region (C) t=100−105 ms. Note that the time window for regions (O) and (C) have been arbitrarily selected between the four possible choices, see Fig. 14. For each region we calculate the detected flux as the integral of the PHS:

  1. 1.

    Φ: total flux integrated in t=0−200 ms

  2. 2.

    Φtn : thermal flux integrated in t=120−160 ms

  3. 3.

    Φp: background flux integrated in t=100−105 ms

As both the detectors are located in the same place, the same incident flux is impinging on both. Therefore, the indirect calculation consists in extracting the sensitivity of the 3He detector by comparing the measured flux with both detectors and the well-known Multi-Blade thermal neutron efficiency [18] and fast neutron sensitivity [31], following the expression 5.

$$ \Phi^{MB}_{i}: \epsilon^{MB}_{i} = \Phi^{He}_{i} : \epsilon^{He}_{i} $$
(5)

It is possible to calculate the 3He efficiency either in the thermal spectrum region or for the background, the index i in equation 5 refers to the two possibilities. The calculation of thermal efficiency serves as a proof of the reliability of this method, whereas the 3He efficiency is widely known [52]. The uncertainties are treated as statistical fluctuations of counts and the error propagation is used for the efficiency calculation. The measurements are performed with the same instrument configuration for both detectors, hence the systematic effects can be considered similar for both.

A further step is needed in order to separate the fast neutron and the γ-ray background contributions. Based on previous studies on the Multi-Blade detector, a threshold discrimination can be applied [18]. It has been proved, indeed, that by setting an appropriate software threshold, it is possible to reject most of the γ-ray contribution from the data. Usually, the γ-rays give rise to less energetic signals, and only few are detected above this value, on the order of 10−7−10−8 [31, 32, 53]. Instead, the fast neutrons have a much broader energy spectrum, which leads to events with higher energy. This contribution can be highlighted by selecting the data above the threshold with a good discrimination to the γ-ray background. We derive the fast neutron flux, Φfn, by the following relation:

$$ \Phi_{fn} = \frac{\Phi-(4*\Phi_{tn})}{4} $$
(6)

Note that the fluxes in equation 6 are calculated after the software threshold discrimination, while the factor 4 is given by the repetition of the regions (O) and (C) in the spectrum (T), as shown in Fig. 14. Moreover, ΦfnΦp when Φp is calculated above threshold. The difference is that Φfn is the average value of background in a time window t=20 ms, while Φp is integrated around the peak of region (C) in an interval t=5 ms.

In order to perform the calculation described above, the Time-of-Flight (ToF) spectrum and the PHS in the three regions of interest (T), (O) and (C) will be presented for both detectors.

In Fig. 15 the ToF spectrum of full time selected t=200 ms is shown, in red for the 3He detector and in blue for the Multi-Blade. This distinction holds to all the subsequent figures. It can be noticed that a much higher background is detected in the case of the 3He-tube compared with the Multi-Blade. Despite the high counting, no saturation effects were detected by either detector.

Fig. 15
figure 15

ToF spectrum obtained with the Multi-Blade detector (blue line) and the 3He-tube (red line) in a time window of 200 ms

For the region (O) (t=120−160 ms) the recorded ToF spectrum reproduces the beam profile, as shown in Fig. 16. The spike at the beginning of this spectrum, and for each of the (O) repetitions shown in Fig. 14 (t=0,80,120 and 160 ms), is significantly higher in the case of the 3He-tube than the one recorded with the Multi-Blade detector by about 2 order of magnitudes. The intensity at the thermal peak (t=128 ms in this case) is higher in the case of the 3He because CRISP is a thermal neutron reflectometer peaked at λ=2.5 Å, where the Multi-Blade has an efficiency of 50 %, with respect to the 85% efficiency of the 3He tube of the instrument [18]. As expected, above 6.5 Å (t=140 ms in Fig. 16), the two detectors efficiency are close enough that they count a similar amount of events.

Fig. 16
figure 16

ToF spectrum obtained with the Multi-Blade detector (blue line) and the 3He-tube (red line) for ToF = 120 - 160 ms

The ToF spectrum in (C) (t=100−105 ms) is shown in Fig. 17 separately for the 3He detector, panel (a), and the Multi-Blade detector, panel (b), because the threshold discrimination is applied individually. The time structure of the pulses is a further indication of the nature of the various background contributions. The gamma flash induces prompt events, indeed, when events below threshold are taken into account, the black pulse in Fig. 17a, it spreads only over 1 ms. The red peak shows, instead, the signals above threshold, this is more intense because, along with the fast neutrons interaction, also a small contribution due to γ-rays of higher energy and spurious scattering from the environment may occur. Note that this peak is broader and extends up to approximately 4 ms. The latter is a further indication that applying the threshold separation it is possible to discriminate most of the γ-rays from the fast neutrons. In the case of the Multi-Blade, Fig. 17b, the difference between the two analysis is still significant, but the counts are much fewer than in the case of the 3He-tube, more than 1 order of magnitude at peak and around 2 orders of magnitude in the integrated region. Not only with the Multi-Blade a lower background is detected, but also for a shorter time, below 1 ms, compared with the ≈ 5 ms recorded with the 3He detector.

Fig. 17
figure 17

a ToF spectrum obtained with the 3He-tube below (black line) and above (red line) threshold. b ToF spectrum obtained with the the Multi-Blade detector below (green line) and above (blue line) threshold. Both are recorded in ToF range between 100 to 105 ms, i.e., when the neutron beam hits against the chopper

In Fig. 18 the Pulse Height Spectrum, normalized over the time, for both detectors is shown in the three regions of study (T) (O) and (C) in panel (a) (b) and (c) respectively. The 3He-detector counts many more background events than the Multi-Blade, as discussed for Fig. 17. The total flux on the detector when the beam passes through the chopper is calculated as the integral of the PHS in Fig. 18a.

Fig. 18
figure 18

a PHS in the full time window 200 ms, b PHS in ToF = 0−200 ms (chopper O) and c PHS in ToF = 100−105 ms

The integral of these PHS is used to perform the calculation of Φfn from equation 6, as described above. The integrated counts for both detectors are listed in Table 1.

Table 1 Integrated counts of PHS for Fig. 18 (a) Φ, (b) Φtn, (c) Φp and the calculated Φfn from equation 6, for both detectors: Multi-Blade (MB) and 3He

Knowing the different Φi for both detectors, the indirect calculation of the 3He efficiency can be derived following equation 5. The sensitivity of the Multi-Blade detector to the fast neutrons is \(\epsilon ^{MB}_{fn} = (1.4 \pm 0.7)\cdot 10^{-5}\). This uncertainty is dominated by the calculation of the solid angle as discussed in section 3, and in reference [31]. Hence the sensitivity of the 3He detector to fast neutrons, \(\epsilon ^{He}_{fn}\) can be expressed as:

$$ \epsilon^{He}_{fn} = \frac{\Phi^{He}_{fn} \cdot \epsilon^{MB}_{fn}}{\Phi^{MB}_{fn}} $$
(7)

By substituting the values in equation 7, a sensitivity \(\epsilon ^{He}_{fn} = (1.2 \pm 0.6) \cdot 10^{-3}\) is obtained. We verify the calculation, considering the Multi-Blade efficiency ≈0.6 at λ=3Å. By equation 5, an efficiency to thermal neutrons of \(\epsilon ^{He}_{tn} = 0.94 \pm 0.09\) is obtained, in agreement with the measured efficiency of the 3He detector, as discussed in [18].

A factor of 2 orders of magnitude between the Multi-Blade and the 3He is obtained in favor of the Multi-Blade. The same sensitivity is expected for other B-10-based gaseous detectors. As demonstrated in [31] and in section 3, for both technologies the main contribution to the fast neutron sensitivity is given by the elastic interaction with the gas. In the case of Multi-Blade, as for most B-10-based gaseous detectors, the gas is Ar/CO2 mixture. These components have a higher atomic mass with respect to the 3He. Therefore, the elastic interaction is more effective in a 3He tube and produce a higher contribution as background signal.

The estimated value is on the order of 10 −3, in agreement with the fast neutron sensitivity measurements performed at the Source Testing Facility discussed in the previous “Fast neutron sensitivity of 3He detector” section.

Conclusions

The signal-to-background ratio strongly affects the figure-of-merit of most of the instruments. Therefore, the background evaluation is of crucial importance to improve the performance of the detector, the efficiency of the instruments, and hence the quality of the measurements together with the data analysis. This paper is focused on the study of fast neutron sensitivity of a 3He-detector. It is the first time this kind of investigation has been performed on a 3He-based thermal neutron detector.

Together with the measurements using a 238Pu/Be fast neutron source, a direct comparison with a boron-based neutron detector, the Multi-Blade detector, has been presented. The aim of this study is to provide a further parameter of discrimination between the new classes of neutron detector technology and the current reference solutions based on the 3He.

The first measurements have been performed at the Source Testing Facility in Lund University in Sweden, while the comparison with the Multi-Blade was part of an experimental test performed at the CRISP reflectometer at ISIS (UK). Two independent methods have been used to obtain the sensitivity: a subtraction method and an indirect calculation. These investigations allow an estimation on the order of magnitude for the sensitivity and are in agreement with each other. Moreover a set of simulations is described in the paper. A good agreement between measurements and simulations is achieved.

For the first set of measurements presented in the paper, the efficiency calculation is affected by an uncertainty no more than a factor two, which is dominated by the systematic effects from the calculation of the solid angle. This leads to a detection efficiency for thermal neutrons of 87%. From the simulations it has been estimated as 94% with an uncertainty of about 20% due to statistical fluctuation. Both values are in agreement with the nominal efficiency of the 3He detector employed for this measurements, which is 96% (at 2.5 Å).

The fast neutron sensitivity has been calculated, as in the case of the Multi-Blade detector, for a fixed software threshold of 100 keV, which is representative of the typical real world applications. We obtained a sensitivity of ≈4·10−3, affected by the systematic uncertainty described before. The value is confirmed by the simulation with a calculated sensitivity of ≈3·10−3 with a 10% uncertainty.

This estimate of the fast neutron sensitivity is also confirmed by the background characterization analysis performed on the campaign of measurements carried out at CRISP. The sensitivity obtained through the indirect calculation applied to the measurements is on the order of 10 −3, while the efficiency to thermal neutrons obtained using the same calculation is 94% with an uncertainty of 10% due to statistical fluctuation. This value is in agreement with the measured efficiency [18].

Along with the good agreement achieved between the different methods used to investigate the fast neutron sensitivity, the direct comparison with Multi-Blade detector shows that this sensitivity is about 2 orders of magnitudes higher for a 3He detector than for a 10B-based neutron detector. Indeed, the fast neutron sensitivity of the Multi-Blade for the fixed software threshold of 100 keV is approximately 10−5, as described in a previous study [31].

The better background discrimination achievable with a boron-based detector has also an impact from a cost effective point of view. The 3He detectors need a lot of shielding material for equivalent performance to prevent from counting fast neutrons. Less shielding can be potentially used with a boron-10-based neutron detector, decreasing the costs, especially for large area applications.

Availability of data and materials

Data essential for evaluation is contained within the figures. Authors are open to placing raw data and processed datasets to aid in replication in a convenient location, but it is currently available by direct request only.

Abbreviations

PHS:

Pulse Height Spectrum

ToF:

Time of Flight

References

  1. Peggs S. ESS Technical Design Report - (ESS-2013-0001). http://eval.esss.lu.se/cgi-bin/public/DocDB/ShowDocument?docid=274.

  2. SNS. http://neutrons.ornl.gov/facilities/SNS/.

  3. J-PARC. http://j-parc.jp/MatLife/en/index.html.

  4. European Spallation Source ESS ERIC. http://europeanspallationsource.se.

  5. Garoby R. The European Spallation Source Design. Physica Scripta. 2018; 93(1):014001.

    Article  ADS  Google Scholar 

  6. Mezei F. New perspectives from new generations of neutron sources. Comptes Rendus Physique. 2007; 8(7):909–920. https://doi.org/10.1016/j.crhy.2007.10.003. Neutron scattering: a comprehensive tool for condensed matter research.

    Article  ADS  Google Scholar 

  7. Vettier C, Carlile CJ, Carlsson P. Progress for the european spallation source in scandinavia. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2009; 600(1):8–9. https://doi.org/10.1016/j.nima.2008.11.097.

    Article  ADS  Google Scholar 

  8. Lindroos M, Bousson S, Calaga R, Danared H, Devanz G, Duperrier R, Eguia J, Eshraqi M, Gammino S, Hahn H, Jansson A, Oyon C, Pape-Moller S, Peggs S, Ponton A, Rathsman K, Ruber R, Satogata T, Trahern G. The European Spallation Source. Nucl Inst Methods Phys Res Sect B Beam Interact Mater Atoms. 2011; 269(24):3258–60. https://doi.org/10.1016/j.nimb.2011.04.012.

    Article  ADS  Google Scholar 

  9. Kouzes RT. The 3He Supply Problem. Tech Rep. 2009.

  10. International collaboration on the development of neutron detectors. http://www.icnd.org.

  11. Zeitelhack K. Search for alternative techniques to helium-3 based detectors for neutron scattering applications. Neutron News. 2012; 23(4):10–3. https://doi.org/10.1080/10448632.2012.725325.

    Article  Google Scholar 

  12. Hurd AJ, Kouzes RT. Why new neutron detector materials must replace helium-3. The European Physical Journal Plus. 2014; 129(10):1–3. https://doi.org/10.1140/epjp/i2014-14236-6.

    Article  Google Scholar 

  13. Anastasopoulos M, Bebb R, Berry K, Birch J, Bryś T, Buffet J-C, Clergeau J-F, Deen PP, Ehlers G, van Esch P, Everett SM, Guerard B, Hall-Wilton R, Herwig K, Hultman L, Höglund C, Iruretagoiena I, Issa F, Jensen J, Khaplanov A, Kirstein O, Higuera IL, Piscitelli F, Robinson L, Schmidt S, Stefanescu I. Multi-Grid detector for neutron spectroscopy: results obtained on time-of-flight spectrometer CNCS. J Instrum. 2017; 12(04):04030.

    Article  Google Scholar 

  14. Birch J, Buffet J-C, Clergeau J-F, Correa J, van Esch P, Ferraton M, Guerard B, Halbwachs J, Hall-Wilton R, Hultman L, Höglund C, Khaplanov A, Koza M, Piscitelli F, Zbiri M. In-beam test of the Boron-10 Multi-Grid neutron detector at the IN6 time-of-flight spectrometer at the ILL. J Phys Conf Ser. 2014; 528(1):012040.

    Article  Google Scholar 

  15. Guerard B, Buffet JC. Ionizing radiation detector - Patent no. 20110215251. Google Patents. Patent App. 13/038. 2011. https://www.google.com/patents/US20110215251.

  16. Birch J, Buffet JC, Correa J, van Esch P, Guérard B, Hall-Wilton R, Höglund C, Hultman L, Khaplanov A, Piscitelli F. 10B4C Multi-Grid as an Alternative to 3He for Large Area Neutron Detectors. IEEE Trans Nuclear Sci. 2013; 60(2):871–78. https://doi.org/10.1109/TNS.2012.2227798.

    Article  ADS  Google Scholar 

  17. Piscitelli F, Buffet JC, Clergeau JF, Cuccaro S, Guérard B, Khaplanov A, Manna QL, Rigal JM, Esch PV. Study of a high spatial resolution 10 B -based thermal neutron detector for application in neutron reflectometry: the Multi-Blade prototype. J Instrum. 2014; 9(03):03007.

    Article  Google Scholar 

  18. Piscitelli F, Mauri G, Messi F, Anastasopoulos M, Arnold T, Glavic A, Höglund C, Ilves T, Higuera IL, Pazmandi P, Raspino D, Robinson L, Schmidt S, Svensson P, Varga D, Hall-Wilton R. Characterization of the Multi-Blade 10B-based detector at the CRISP reflectometer at ISIS for neutron reflectometry at ESS. J Instrum. 2018; 13(05):05009.

    Article  Google Scholar 

  19. Mauri G, Messi F, Anastasopoulos M, Arnold T, Glavic A, Höglund C, Ilves T, Lopez Higuera I, Pazmandi P, Raspino D, Robinson L, Schmidt S, Svensson P, Varga D, Hall-Wilton R, Piscitelli F. Neutron reflectometry with the Multi-Blade 10B-based detector. Proc R Soc Lond A Math Phys Eng Sci. 2018; 474(2216). https://doi.org/10.1098/rspa.2018.0266.

    Article  ADS  Google Scholar 

  20. Henske M, Klein M, Köhli M, Lennert P, Modzel G, Schmidt CJ, Schmidt U. The 10B based Jalousie neutron detector – An alternative for 3He filled position sensitive counter tubes. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2012; 686:151–5. https://doi.org/10.1016/j.nima.2012.05.075.

    Article  ADS  Google Scholar 

  21. Croci G, et al.Diffraction measurements with a boron-based GEM neutron detector. EPL. 2014; 107(1):12001. https://doi.org/10.1209/0295-5075/107/12001.

    Article  ADS  Google Scholar 

  22. Albani G, Cippo EP, Croci G, Muraro A, Schooneveld E, Scherillo A, Hall-Wilton R, Kanaki K, Höglund C, Hultman L, Birch J, Claps G, Murtas F, Rebai M, Tardocchi M, Gorini G. Evolution in boron-based GEM detectors for diffraction measurements: from planar to 3D converters. Meas Sci Technol. 2016; 27(11):115902.

    Article  ADS  Google Scholar 

  23. Lacy JL, Athanasiades A, Sun L, Martin CS, Lyons TD, Foss MA, Haygood HB. Boron-coated straws as a replacement for 3He-based neutron detectors. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2011; 652(1):359–63. https://doi.org/10.1016/j.nima.2010.09.011. Symposium on Radiation Measurements and Applications (SORMA) {XII} 2010.

    Article  ADS  Google Scholar 

  24. Köhli M, Allmendinger F, Häußler W, Schröder T, Klein M, Meven M, Schmidt U. Efficiency and spatial resolution of the CASCADE thermal neutron detector. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2016; 828:242–9. https://doi.org/10.1016/j.nima.2016.05.014.

    Article  ADS  Google Scholar 

  25. Pfeiffer D, Resnati F, Birch J, Etxegarai M, Hall-Wilton R, Höglund C, Hultman L, Llamas-Jansa I, Oliveri E, Oksanen E, Robinson L, Ropelewski L, Schmidt S, Streli C, Thuiner P. First measurements with new high-resolution gadolinium-GEM neutron detectors. J Instrum. 2016; 11(05):05011.

    Article  Google Scholar 

  26. Pfeiffer D, Resnati F, Birch J, Hall-Wilton R, Höglund C, Hultman L, Iakovidis G, Oliveri E, Oksanen E, Ropelewski L, Thuiner P. The μTPC method: improving the position resolution of neutron detectors based on MPGDs. J Instrum. 2015; 10(04):04004.

    Article  Google Scholar 

  27. Mireshghi A, Cho G, Drewery J, Jing T, Kaplan SN, Perez-Mendez V, Wildermuth D. Amorphous silicon position sensitive neutron detector. IEEE Trans Nuclear Sci. 1992; 39(4):635–40. https://doi.org/10.1109/23.159678.

    Article  ADS  Google Scholar 

  28. Schulte RL, Swanson F, Kesselman M. The use of large area silicon sensors for thermal neutron detection. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 1994; 353(1):123–7. https://doi.org/10.1016/0168-9002(94)91617-9.

    Article  ADS  Google Scholar 

  29. Petrillo C, Sacchetti F, Toker O, Rhodes NJ. Solid state neutron detectors. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 1996; 378(3):541–51. https://doi.org/10.1016/0168-9002(96)00487-1.

    Article  ADS  Google Scholar 

  30. Mauri G, Mariotti M, Casinini F, Sacchetti F, Petrillo C. Development of pulse shape analysis for noise reduction in Si-based neutron detectors. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2018; 910:184–93. https://doi.org/10.1016/j.nima.2018.09.071.

    Article  ADS  Google Scholar 

  31. Mauri G, Messi F, Kanaki K, Hall-Wilton R, Karnickis E, Khaplanov A, Piscitelli F. Fast neutron sensitivity of neutron detectors based on boron-10 converter layers. J Instrum. 2018; 13(03):03004–171205614.

    Article  Google Scholar 

  32. Piscitelli F, Messi F, Anastasopoulos M, Bryś T, Chicken F, Dian E, Fuzi J, Höglund C, Kiss G, Orban J, Pazmandi P, Robinson L, Rosta L, Schmidt S, Varga D, Zsiros T, Hall-Wilton R. The Multi-Blade Boron-10-based neutron detector for high intensity neutron reflectometry at ESS. J Instrum. 2017; 12(03):03013.

    Article  Google Scholar 

  33. CRISP instrument manual 2010. 2010. https://www.isis.stfc.ac.uk/Pages/crisp-instrument-manual-nov-2010.pdf.

  34. ISIS Neutron and Muon Source. https://www.isis.stfc.ac.uk.

  35. Messi F, Perrey H, Fissum K, Akkawi M, Jebali RA, Annand JRM, Bentley P, Boyd L, Cooper-Jensen CP, DiJulio DD, Freita-Ramos J, Hall-Wilton R, Huusko A, Ilves T, Issa F, Jalgen A, Kanaki K, Karnickis E, Khaplanov A, Koufigar S, Maulerova V, Mauri G, Mauritzson N, Pei W, Piscitelli F, Rofors E, Scherzinger J, Soderhielm H, Soderstrom D, Stefanescu I. The neutron tagging facility at Lund University. arXiv:1711.10286 (submitted to IAEA Technical Report on Modern Neutron Detection 2017). 2017.

  36. General Electric (GE) Reuter Stokes. https://www.industrial.ai.

  37. Scherzinger J, Annand JRM, Davatz G, Fissum KG, Gendotti U, Hall-Wilton R, Håkansson E, Jebali R, Kanaki K, Lundin M, Nilsson B, Rosborge A, Svensson H. Tagging fast neutrons from an 241Am/9Be source. Appl Radiat Isot. 2015; 98:74–9. https://doi.org/10.1016/j.apradiso.2015.01.003.

    Article  Google Scholar 

  38. Scherzinger J, Jebali RA, Annand JRM, Fissum KG, Hall-Wilton R, Koufigar S, Mauritzson N, Messi F, Perrey H, Rofors E. A comparison of untagged gamma-ray and tagged-neutron yields from 241AmBe and 238PuBe sources. Appl Radiat Isot. 2017; 127:98–102. https://doi.org/10.1016/j.apradiso.2017.05.014.

    Article  Google Scholar 

  39. Rossi E. Characterisation of the Spatial Resolution and the Gamma-ray Discrimination of Helium-3 Proportional Counters. Master’s thesis. 2017. http://arxiv.org/abs/1702.06501.

  40. Tsutomu IIJIMA and Takehiko MUKAIYAMA and Keisho SHIRAKATA. Measurement of Fast Neutron Spectrum with Helium-3 Proportional Counter. J Nucl Sci Technol. 1971; 8(4):192–200. https://doi.org/10.1080/18811248.1971.9735315. https://www.tandfonline.com/doi/pdf/10.1080/18811248.1971.9735315.

    Article  Google Scholar 

  41. Swiderski L, Moszynski M, Wolski D, Iwanowska J, Szczesniak T, Pausch G, Plettner C, Stein J, Schotanus P, Hurlbut C. Comparison of neutron detection efficiency using a he-3 counter and a boron-10 loaded liquid scintillator ej309b5. In: 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC): 2009. p. 892–7. https://doi.org/10.1109/NSSMIC.2009.5401557.

  42. Swiderski L, Moszynski M, Wolski D, Batsch T, Iwanowska J, Nassalski A, Syntfeld-Kazuch A, Szczesniak T, Kniest F, Kusner MR, Pausch G, Stein J, Klamra W, Schotanus P, Hurlbut C. Further Study of Boron-10 Loaded Liquid Scintillators for Detection of Fast and Thermal Neutrons. IEEE Trans Nuclear Sci. 2010; 57(1):375–80. https://doi.org/10.1109/TNS.2009.2037223.

    Article  ADS  Google Scholar 

  43. Pino F, Stevanato L, Cester D, Nebbia G, Sajo-Bohus L, Viesti G. Detecting fast and thermal neutrons with a boron loaded liquid scintillator, EJ-339A. Appl Radiat Isot Incl Data Instrum Methods Agric Ind Med. 2014; 92:6–11. https://doi.org/10.1016/j.apradiso.2014.05.025.

    Google Scholar 

  44. NIST National Nuclear Data Center. https://www.nndc.bnl.gov.

  45. Agostinelli S, Allison J, et al.Geant4 - A Simulation Toolkit. Nucl Instrum Meth A. 2003; 506:250–303.

    Article  ADS  Google Scholar 

  46. Allison J, et al.Geant4 developments and applications. IEEE Trans Nucl Sci. 2006; 53:07.

    Article  Google Scholar 

  47. Allison J, Amako K, et al.Recent Developments in Geant4. Nucl Instrum Meth A. 2016; 835:186–225.

    Article  ADS  Google Scholar 

  48. Kittelmann T, et al.Geant4 Based Simulations for Novel Neutron Detector Development. J Phys Conf Ser. 2014; 513:022017.

    Article  Google Scholar 

  49. Kittelmann T, Boin M. Polycrystalline neutron scattering for Geant4: NXSG4. Comput Phys Commun. 2015; 189(0):114–8. https://doi.org/10.1016/j.cpc.2014.11.009.

    Article  ADS  Google Scholar 

  50. Kittelmann T, Kanaki K, Klinkby E, Cai XX, Cooper-Jensen CP, Hall-Wilton R. Using backscattering to enhance efficiency in neutron detectors. IEEE TNS. 2017; 64(6):1562–73. https://doi.org/10.1109/TNS.2017.2695404.

    Google Scholar 

  51. Deiev OS. GEANT 4 simulation of neutron transport and scattering in media (3-85/60). Voprosy Atomnoj Nauki i Tekhniki. 2013; 46(11):236–241.

    Google Scholar 

  52. Mühlbauer S, Heinemann A, Wilhelm A, Karge L, Ostermann A, Defendi I, Schreyer A, Petry W, Gilles R. The new small-angle neutron scattering instrument SANS-1 at MLZ—characterization and first results. Nucl Inst Methods Phys Res Sect A Accelerators Spectrometers Detectors Assoc Equip. 2016; 832:297–305. https://doi.org/10.1016/j.nima.2016.06.105.

    Article  ADS  Google Scholar 

  53. Khaplanov A, Piscitelli F, Buffet J-C, Clergeau J-F, Correa J, van Esch P, Ferraton M, Guerard B, Hall-Wilton R. Investigation of gamma-ray sensitivity of neutron detectors based on thin converter films. J Instrum. 2013; 8(10):10025.

    Article  Google Scholar 

Download references

Acknowledgements

This work was carried out as a part of the collaboration between the European Spallation Source (ESS - Sweden), the Lund University (LU - Sweden) and the University of Perugia (Italy).

The work was carried out at the Source Testing Facility, Lund University (LU - Sweden). Computing resources provided by DMSC Computing Centre:https://europeanspallationsource.se/data-management-software/computing-centre.

The authors would like to thank the ISIS detector group for the support during the tests.

The authors thank the CRISP instrument scientists R. Dalgliesh and C. Kinane for providing the beam time and the instrument support necessary for this detector test.

Funding

This work was supported by the BrightnESS project (Horizon 2020, INFRADEV-3-2015, grant number 676548)

Author information

Authors and Affiliations

Authors

Contributions

GM drafted the manuscript. GM, FP and FM conceived the experiment. GM analysed the data. KK performed the simulation. All the authors contributed to the interpretation of the results. All the authors gave their final approval for publication.

Corresponding author

Correspondence to Giacomo Mauri.

Ethics declarations

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mauri, G., Messi, F., Kanaki, K. et al. Fast neutron sensitivity for 3He detectors and comparison with Boron-10 based neutron detectors. EPJ Techn Instrum 6, 3 (2019). https://doi.org/10.1140/epjti/s40485-019-0052-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1140/epjti/s40485-019-0052-x

Keywords